Features
- Easy keypad operation with LCD display
- Wavelengths range is 330-800nm, 5 standard filters, 3 more filters optional
- Reagent open system supports flow cell and cuvette mode, close system on request
- Analytical models include kinetic, fixed time, bichromatic end point with or without reagent blank or sample blank, linear or non-linear calibration curves
- Up to 60 programs and 2200 test results can be stored
- Excellent QC function, 2 controls per test
- Internal thermal-sensitive printer
- Compact design enables USER-DIY maintenance
- Multi-language software available on request
High Stability & Accuracy
l Use the most advanced binuclear embedded digital circuit system, to make the analytical ability great and stable
l 2 counting channels of WBC and RBC, to make the test results more accurate and less cross- contamination
l Use CPLD technology in data processing, to avoid the interference of electromagnetic signals
Reliable Operation System
l ARM intelligent micro-processing technology
l Data analysis: intelligent pulse formulation technology
l Classification: floating landmark technology
l Module design, fast maintenance
User-friendly System
l 10.4''large color LCD displays main menu and all parameters & color histograms in one screen
l Brief operation instruction built-in software, easy for user to view
l Separate module design, very simple mechanical structure, easy maintenance
l Self check for hardware, alarm for error
l Reagent depletion alarm for Lyse, Cleanser and Diluent, as optional function
Technical Features
l 2 counting mode: whole blood, pre-dilution blood
l Sample pre-dilution: internal diluter, add 1ml diluent automatically by the needle, no need to add diluent manually to avoid deviation.
l Calibration: automated or manual
l QC: comprehensive QC program, forming and printing the histogram automatically
l Block removal: Real time monitor, high ignition, soak, flush and back flush
- Absorbance Range:-0.500-3.500 Abs
- Resolution: 0.001Abs (displayed), 0.0001Abs (calculated)
- Light Source: Halogen lamp
- Wavelengths: 340,405,500,546,620nm, 3 more filters optional
- Wavelength Accuracy: ±2nm
- Band Width: ≤10nm
- Cross Contamination Rate: ≤1%
- Interface: RS-232 serial cable
- Flow Cell: Metal - quartz flow cell
- Temperature Control: 25°C, 30°C, 37°C; ±0.5°C and ambient temperature
- Display: LCD display
- Output: Internal printer
- Power Supply: AC 110V/220V ± 10%, 50HZ/60Hz
- Net Weight: 7KG
- Dimensions L x W x H (mm): 360 x 318 x 160